Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 22
Filtrar
1.
Rev. bras. ortop ; 57(6): 968-974, Nov.-Dec. 2022. tab, graf
Artigo em Inglês | LILACS | ID: biblio-1423630

RESUMO

Abstract Objective The present study aims to evaluate the influence of different positioning of the hip femoral prosthesis on the stress and strain over this implant. Methods A femoral prosthesis (Taper - Víncula, Rio Claro, SP, Brazil) was submitted to a stress and strain analysis using the finite element method (FEM) according to the International Organization for Standardization (ISO) 7206-6 Implants for surgery - Partial and total hip joint prostheses - Part 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components standard. The analysis proposed a branch of the physical test with a +/− 5° angle variation on the standard proposed for α and β variables. Results The isolated +/− 5° variation on the α angle, as well as the association of +/− 5° variation on the α and β angles, presented significant statistical differences compared with the control strain (p= 0.027 and 0.021, respectively). Variation on angle β alone did not result in a significant change in the strain of the prosthesis (p= 0.128). The stem positioning with greatest implant strain was α = 5° and β = 14° (p= 0.032). Conclusion A variation on the positioning of the prosthetic femoral stem by +/− 5° in the coronal plane and/or the association of a +/− 5° angle in coronal and sagittal planes significantly influenced implant strain.


Resumo Objetivo Avaliar a influência da variação do posicionamento da prótese femoral do quadril na tensão e na deformação produzidas neste implante. Métodos Utilizou-se a análise de tensão e de deformação da prótese femoral (Taper, Víncula, Rio Claro, SP, Brasil) pelo método de elementos finitos (MEF) de acordo com a norma ISO 7206-6 Implants for surgery - Partial and total hip joint prostheses-Part 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components. A análise propôs uma ramificação do ensaio físico, com variação da angulação de +/− 5° sobre a proposta normativa das variáveis α e β. Resultados Ao comparar com a deformação controle, houve significância estatística com a angulação isolada de +/− 5° do ângulo α, bem como com a associação de +/− 5° nas angulações α e β (p= 0,027 e 0,021, respectivamente). Já com a variação apenas do ângulo β, não houve variação significativa na deformação da prótese (p= 0,128). A posição da haste com maior deformação no implante foi com α = 5° e β = 14° (p= 0,032). Conclusão A variabilidade de posicionamento da haste femoral protética de +/− 5° no plano coronal e/ou a associação da angulação de +/− 5° nos planos coronal e sagital interferiu de forma significativa na deformação do implante.


Assuntos
Humanos , Análise de Falha de Equipamento , Artroplastia de Quadril/efeitos adversos , Análise de Elementos Finitos , Prótese de Quadril
2.
Acta ortop. bras ; 30(5): e255298, 2022. tab
Artigo em Inglês | LILACS-Express | LILACS | ID: biblio-1403042

RESUMO

ABSTRACT Objective: To evaluate the impact of the COVID-19 pandemic on hospital admission and mortality indicators in older adults with fractures of the proximal femur. Methods: Observational and retrospective study that took place from June 2016 to 2020. Patients of both genders who underwent surgical treatment for fractures of the proximal end of the femur, aged over 60 years, were included. Results: The population consisted of 379 patients, treated before (group 1; N = 278; 73.35%) and during the pandemic (group 2; N = 101; 26.65%). Higher mortality was observed in group 2 (N = 24; 23.8%) versus group 1 (N = 10; 3.6%), p < 0.001. The highest proportion of deaths in group 2 was maintained in patients aged 70-79 years (p = 0.011), 80-89 years (p ≤ 0.001) and > 90 years (p ≤ 0.001). In addition, the preoperative time and hospital stay were longer in group 2 compared to group 1 (p ≤ 0.001). Conclusion: The present study demonstrated that the pandemic period increased the mortality rate and the preoperative and hospitalization time in older patients with femur fractures. Thus, the pandemic has affected the care of fractures of the proximal femur in older adults, which reinforces the need to adopt measures to reduce complications and mortality. Level of Evidence II, Retrospective Study.


RESUMO Objetivo: Avaliar o impacto da pandemia de COVID-19 nos indicadores de internação e mortalidade hospitalar em idosos com fraturas do fêmur proximal. Métodos: Estudo observacional e retrospectivo, que ocorreu de junho de 2016 a 2020. Foram incluídos pacientes de ambos os sexos submetidos ao tratamento cirúrgico para fratura da extremidade proximal do fêmur, com idade acima de 60 anos. Resultados: A população foi composta por 379 pacientes, atendidos antes (grupo 1; N = 278; 73,35%) e durante a pandemia (grupo 2; N = 101; 26,65%). Foi verificada maior mortalidade no grupo 2 (N = 24; 23,8%) versus o grupo 1 (N = 10; 3,6%), p < 0,001. A maior proporção de óbitos no grupo 2 se manteve nos pacientes com 70-79 anos (p = 0,011), 80-89 anos (p ≤ 0,001) e > 90 anos (p ≤ 0,001). Os períodos pré-operatório e de permanência hospitalar foram maiores no grupo 2 em comparação ao grupo 1 (p ≤ 0,001). Conclusão: Este estudo demonstrou que o período pandêmico aumentou a taxa de mortalidade e o tempo pré-operatório e de internação em pacientes idosos com fratura do fêmur. Dessa forma, a pandemia tem afetado o atendimento das fraturas do fêmur proximal em idosos, o que reforça a necessidade da adoção de medidas de redução de complicações e mortalidade. Nível de Evidência II, Estudo Retrospectivo.

3.
Acta ortop. bras ; 29(6): 293-296, Nov.-Dec. 2021. tab, graf
Artigo em Inglês | LILACS-Express | LILACS | ID: biblio-1349898

RESUMO

ABSTRACT Objective: To determine the efficacy of Deep Venous Thrombosis (DVT) prophylaxis protocol in patients undergoing total hip arthroplasty (THA), and to verify differences in the rates of this complication when comparing primary replacement surgeries with revision surgeries. Methods: We conducted a retrospective study of patients operated between 2012 and 2018, with inclusion and non-inclusion criteria determined by the researchers. A protocol was created to compile data including 92 patients, amounting to 110 operated hips, divided by gender, age, laterality, among other criteria. For THA cases, low molecular weight heparin chemoprophylaxis was used, associated with the use of pneumatic and elastic compression, concomitant to deambulation as soon as possible. The detection of DVT was determined by clinical evaluation and imaging exams such as: simple radiographs, ultrasound, arterial and venous color doppler, and laboratory tests. Results: The use of the protocol was very effective in our study. Only one (1.09%) case of deep venous thrombosis was found. Conclusion: The use of thromboprophylaxis for DVT is indispensable, as was clearly demonstrated by these cases. We observed only one (1.09%) case of DVT in our series. Level of Evidence III, Retrospective Study.


RESUMO Objetivo: Determinar a eficácia de protocolo de profilaxia contra a Trombose Venosa Profunda (TVP) em pacientes submetidos à artroplastia total do quadril (ATQ) e verificar diferenças nas taxas desta complicação ao comparar as cirurgias de substituição primária com as de revisão. Métodos: Realizamos um estudo retrospectivo de pacientes operados entre 2012 e 2018, com critérios de inclusão e não inclusão determinados pelos pesquisadores. Foi criado um protocolo para a compilação de dados no qual foram incluídos 92 pacientes, estes com 110 quadris operados divididos por sexo, idade, lateralidade, entre outros critérios. Para os casos de ATQ, utilizamos a quimioprofilaxia com heparina de baixo peso molecular associada ao uso de compressão pneumática e elástica, concomitante à deambulação assim que possível. A detecção da TVP foi determinada pela avaliação clínica e por exames de imagem como: radiografias simples, exame ultrassonográfico, doppler colorido arterial e venoso e exames laboratoriais. Resultados: A utilização do protocolo se mostrou bastante eficaz em nosso estudo. Foi encontrado apenas um (1,09%) caso de trombose venosa profunda. Conclusão: A utilização de tromboprofilaxia para a TVP é indispensável e consagrada nestes casos. Observamos apenas um (1,09%) caso de TVP em nossa casuística. Nível de Evidência III, Estudo Retrospectivo.

4.
Acta ortop. bras ; 29(5): 246-248, Sept.-Oct. 2021. tab, graf
Artigo em Inglês | LILACS-Express | LILACS | ID: biblio-1339065

RESUMO

ABSTRACT Objective: To date, the literature lacks consensus on the most efficient method to measure the range of motion of an in vitro prosthetic system. In this study, we propose the use of a relatively low-cost online software to measure the range of motion of hip prosthetic implants manufactured in Brazil and compare its results with the current technical standards for hip arthroplasty. Methods: Three different diameters of femoral heads were evaluated (28 mm, 32 mm, and 36 mm). The mean values of the angular displacement of the prosthesis in each motion axis were obtained by computer simulations. Results: The range of motion with each femoral head was 28mm (extension/flexion: 148°, internal/external rotation: 179°, adduction/abduction: 107°), 32 mm (152°/185°/114°), and 36 mm (158°/193°/120°). Conclusion: The computational method showed that the larger the femoral head, the greater the range of motion of the hip joint prosthetic system. Additional clinical studies are necessary to compare the physical results obtained with the values found in this study by computational modeling. Level of evidence V, Experimental study.


RESUMO Objetivo: A maneira mais eficiente de se aferir a amplitude de movimento de um sistema protético in vitro ainda não é bem estabelecido na literatura. Assim, o presente estudo propõe a utilização de um software online de custo relativamente baixo para mensuração da amplitude de movimento de um conjunto protético nacional de quadril e comparar os resultados obtidos com a norma técnica vigente para as artroplastias de quadril. Métodos: A avaliação foi realizada com três diferentes diâmetros de cabeças femorais (28 mm, 32 mm e 36 mm); os valores médios do deslocamento angular da prótese em cada eixo de movimento foram obtidos por meio de simulações computacionais no programa Autodesk Inventor. Resultados: as amplitudes de movimento obtidas foram: cabeça 28mm (extensão/flexão: 148°, rotação interna/externa: 179°, adução/abdução: 107°), cabeça 32° (152°/185°/114°), cabeça 36° (158°/193°/120°). Conclusão: O método computacional utilizado no presente estudo possibilitou concluir que quanto maior a cabeça femoral, maior será a amplitude de movimento do sistema protético para articulação de quadril. Novos estudos clínicos, tanto pré quanto pós-operatórios, devem ser realizados para comparar os resultados físicos obtidos com os valores encontrados nesta avaliação por meio de modelo computacional. Nível de Evidência V, Estudo experimental.

5.
Rev. bras. ortop ; 55(5): 597-604, Sept.-Oct. 2020. tab, graf
Artigo em Inglês | LILACS | ID: biblio-1144213

RESUMO

Abstract Objective This study aims to compare the in vitro wear rate of crosslinked, high molecular weight polyethylene coupled to 36-mm diameter ceramic heads and 32-mm diameter metal heads. Methods Ceramic-on-polyethylene (36 mm) and metal-on-polyethylene (32 mm) tribological pairs were submitted to biomechanical tests in a simulator to determine the wear rate after 15 × 106 cycles. Results A statistically significant difference (p= 0.0005) was detected when comparing the wear rate of assemblies with metallic heads (average wear: 14.12 mg/MC) and ceramic heads (average wear: 7.46 mg/MC). Conclusion The present study demonstrated the lower wear rate in prosthetic assemblies using 36-mm crosslinked ceramic-on-polyethylene tribological pairs compared to 32-mm crosslinked metal-on-polyethylene assemblies. This finding demonstrates the effectiveness of ceramic-on-polyethylene tribological pairs, even with large diameter heads.


Resumo Objetivo: O objetivo do presente estudo foi comparar, in vitro, a taxa de desgaste do polietileno de alto peso molecular reticulado acoplado a cabeças cerâmicas de 36 mm de diâmetro e acoplado a cabeças metálicas de 32 mm de diâmetro. Métodos: Foram realizados ensaios biomecânicos em simulador de desgaste para os pares tribológicos cerâmica-poli (36 mm) e metal-poli (32 mm) a fim de verificar a taxa de desgaste após em 15 × 106 ciclos. Resultados: Na comparação entre as medidas de taxa de desgaste dos conjuntos com cabeças metálicas (média:14,12 mg/MC) e cerâmicas (média:7,46 mg/MC) houve diferença estatitsticamente significativa (p= 0,0005). Conclusão: O presente estudo demonstrou menor taxa de desgaste em conjuntos protéticos que utilizaram o par tribológico cerâmica-polietileno reticulado de 36 mm em comparação aos conjuntos com metal-polietileno reticulado de 32 mm. Tal achado demonstra a eficácia do par tribológico cerâmica-poli, mesmo com a utilização de cabeças de grande diâmetro.


Assuntos
Desenho de Prótese , Pesos e Medidas , Técnicas In Vitro , Cerâmica , Artroplastia de Quadril , Polietileno , Quadril
6.
Rev. bras. ortop ; 52(supl.1): 3-13, 2017. tab, graf
Artigo em Inglês | LILACS | ID: biblio-1042412

RESUMO

Abstract While the value of national arthroplasty registries (NAR) for quality improvement in total hip arthroplasty (THA) has already been widely reported, some methodological limitations associated with observational epidemiological studies that may interfere with the assessment of safety and efficacy of prosthetic implants have recently been described in the literature. Among the main limitations of NAR, the need for at least 80% compliance of all health institutions covered by the registry is emphasized; completeness equal or greater than 90% of all THA performed; restricted data collection; use of revision surgery as the sole criterion for outcome; and the inability of establishing a definite causal link with prosthetic dysfunction. The present article evaluates the advantages and limitations of NAR, in the light of current knowledge, which point to the need for a broader data collection and the use of more structured criteria for defining outcomes. In this scenario, the authors describe of idealization, conceptual and operational structure, and the project of implantation and implementation of a multicenter registry model, called Rempro-SBQ, which includes healthcare institutions already linked to the Brazilian Hip Society (Sociedade Brasileira de Quadril [SBQ]). This partnership enables the collection of more reliable and comprehensive data at a higher hierarchical level, with a significant reduction in maintenance and financing costs. The quality improvement actions supported by SBQ may enhance its effectiveness and stimulate greater adherence for collecting, storing, interpreting, and disseminating information (feedback).


Resumo Ainda que os registros nacionais de artroplastias (RNAs) sejam um importante instrumento de controle e melhoria da qualidade da artroplastia total de quadril (ATQ), algumas limitações metodológicas associadas aos estudos epidemiológicos observacionais, que podem interferir na avaliação da segurança e eficácia dos implantes protéticos, têm sido recentemente descritas na literatura.Dentre as principais limitações destacam-se a necessidade de cobertura mínima de 80% das instituições hospitalares da região objeto do registro; integralidade mínima de 90% de todas as ATQs feitas; coleta de informações mais restritas; uso da cirurgia de revisão como critério único de desfecho e a dificuldade de se estabelecer um nexo causal com a disfunção protética.No presente artigo avaliamos as vantagens e limitações dos RNAs, à luz dos conhecimentos atuais, que apontam para a necessidade da coleta de informações mais amplas e de uso de critérios mais estruturados na definição de desfechos.Nesse cenário, descrevemos os processos de idealização, a estrutura conceitual e operacional e o projeto de implantação e implementação de um modelo de registro multicêntrico, denominado Rempro-SBQ, que inclui os centros hospitalares de treinamento em cirurgia de quadril já ligados institucionalmente à Sociedade Brasileira de Quadril (SBQ). Essa parceria possibilita, simultaneamente, a coleta de informações mais abrangentes e de elevado nível hierárquico, de forma confiável, com redução bastante significativa dos custos de manutenção e financiamento. As ações de melhoria da qualidade, amparadas pela SBQ, podem protagonizar uma condição de maior efetividade e maior adesão aos procedimentos de coleta, armazenamento, interpretação e divulgação das informações.


Assuntos
Falha de Prótese , Sistema de Registros , Artroplastia de Quadril
7.
Acta ortop. bras ; 24(6): 327-329, Nov.-Dec. 2016. tab, graf
Artigo em Português | LILACS | ID: biblio-827693

RESUMO

ABSTRACT Objective: To assess the adequacy to the Brazilian population of orthopedic implants used for treatment of proximal femoral fractures. Methods: The neck-shaft angle of the femur of 101 patients was measured in anteroposterior pelvis radiographs and these measurements were correlated to gender, age, height, weight and ethnicity. In addition, we compared the values of the neck -shaft angle with the angulation of the main implants available in the Brazilian market for the treatment of transtrochanteric fractures. Results: Of the 101 measurements, an average of 130.9±6.7° was obtained, ranging from 112° to 150°. Correlating these measurements with epidemiological variables, only age was statistically significant. Conclusion: Most of the analyzed population presented anatomical characteristics that allow the proper use of these implants to treat transtrochanteric fractures, as indicated from the analysis of neck-shaft angles. Nonetheless, 4% of individuals did not fit this pattern and would have required alternative implants. Level of Evidence III, Study of nonconsecutive patients; without consistently applied reference ''gold'' standard.

8.
Rev. bras. ortop ; 48(4): 374-376, ago. 2013. graf
Artigo em Inglês | LILACS | ID: lil-690269

RESUMO

OBJECTIVE: The overuse injuries in the hip joint occur commonly in sports practitioners and currently due to technical advances in diagnostic imaging, especially magnetic resonance imaging (MRI), are often misdiagnosed. Recently, a group of people were reported, all female, with pain and swelling in the pelvic region.T2-weighted MRI showed increased signal in the enthesis of the iliotibial band (ITB) along the lower border of the iliac tubercle. We report a case of a 34 year old woman, non-professional runner, with pain at the iliac crest with no history of trauma and whose MRI was compatible with the proximal iliotibial band syndrome. .


As lesões por sobrecarga na articulação do quadril ocorrem comumente em praticantes de esporte e atualmente, por causa do avanço das técnicas de diagnóstico por imagem, especialmente a ressonância magnética (RM), são frequentemente diagnosticadas. Recentemente, foi estudado um grupo de pacientes, todos do sexo feminino, com quadro de dor e edema na região do tubérculo ilíaco. A RM ponderada em T2 demonstrava aumento de sinal na êntese da banda iliotibial (BIT) ao longo da margem inferior do tubérculo ilíaco. Relatamos um caso de uma mulher de 34 anos, corredora não profissional, com quadro de dor na crista ilíaca sem história de trauma e cuja RM era compatível com a síndrome da banda iliotibial proximal. .


Assuntos
Humanos , Feminino , Adulto Jovem , Quadril , Ílio , Imageamento por Ressonância Magnética , Dor
9.
Rev. bras. ortop ; 47(2): 165-168, mar.-abr. 2012. ilus, tab
Artigo em Português | LILACS | ID: lil-643092

RESUMO

OBJETIVOS: Avaliar a eficácia da osteossíntese de fraturas do colo femoral com dois em vez de três parafusos. MÉTODOS: Avaliadas, retrospectivamente, 39 fraturas, divididas em grupos nos quais foram utilizados dois (n = 28) parafusos paralelos e três parafusos (n = 11) na configuração de triângulo invertido segundo a técnica AO. Os pacientes foram acompanhados até o desfecho: consolidação ou falha. RESULTADOS: No grupo em que se utilizou dois parafusos, observamos consolidação em 23 das 28 fraturas (82%). No grupo de três parafusos observou-se consolidação em seis das 11 fraturas (55%). Não houve diferença estatística entre os valores obtidos. CONCLUSÃO: Não houve diferença no prognóstico dessas fraturas quando tratadas com dois parafusos paralelos ou três parafusos em triângulo invertido segundo a técnica AO-ASIF. Mais estudos são necessários para estabelecer conclusão definitiva.


OBJECTIVES: To evaluate the efficacy of osteosynthesis on femoral neck fractures using two instead of three screws. METHODS: Thirtynine fractures were retrospectively evaluated, divided into groups in which two screws were used in parallel (n = 28) or three screws (n =11) in an inverted triangle configuration (in accordance with the AO technique). The patients were then followed up until reaching the outcome of either consolidation or failure. RESULTS: In the group in which two screws were used, consolidation was observed in 23 of the 28 fractures (82%). In the group in which three screws were used, consolidation was observed in 6 of the 11 fractures (55%). There was no statistically significant difference between these percentages. CONCLUSION: There was no difference in the prognosis for these fractures when treated using two screws in parallel or three screws in an inverted triangle in accordance with the AO technique. Further studies are needed in order to establish a definitive conclusion.


Assuntos
Humanos , Masculino , Feminino , Adolescente , Adulto , Pessoa de Meia-Idade , Idoso de 80 Anos ou mais , Parafusos Ósseos , Fraturas do Colo Femoral/reabilitação , Fraturas do Colo Femoral/terapia
10.
Rev. bras. ortop ; 47(6): 780-783, 2012. ilus
Artigo em Português | LILACS | ID: lil-666227

RESUMO

A associação entre dor no quadril, anormalidade do músculo quadrado femoral na ressonância magnética (RM) e estreitamento do espaço isquiofemoral tem sido relatada na literatura atual, levantando a possibilidade de que essa lesão muscular seja causada pelo impacto isquiofemoral. Tal diagnóstico foi observado em algumas mulheres de meia-idade com ou sem história de trauma ou cirurgia. Os autores relatam aqui uma mulher de 31 anos de idade, que apresentava dor no quadril sem história de trauma, com evidência de estreitamento do espaço isquiofemoral e edema no músculo quadrado femoral. Foi realizado tratamento não cirúrgico com alívio dos sintomas. O diagnóstico de impacto isquiofemoral deve ser considerado em pacientes do sexo feminino com queixa de dor no quadril sem outra causa evidente.


An association between hip pain and quadratus femoris muscle abnormality on magnetic resonance imaging (MRI) with concurrent narrowing of the ischiofemoral space has been reported in the recent literature. This raises the possibility that the muscle damage observed is due to ischiofemoral impingement. This diagnosis has been noted in middle-aged females either with or without a history of trauma or surgery, is a rarely described feature. We report here on a 31-year-old woman who presented with non-traumatic hip pain and evidence of narrowing of the ischiofemoral space and edema in the quadratus femoris. Nonsurgical treatment was administered, which relieved her hip pain. The diagnosis of ischiofemoral impingement should be considered in female patients complaining of hip pain without any other evident cause.


Assuntos
Adulto , Ísquio/patologia , Nervo Isquiático/fisiopatologia , Nervo Isquiático/patologia , Dor , Quadril/patologia
11.
Acta ortop. bras ; 15(4): 214-217, 2007. tab
Artigo em Português | LILACS | ID: lil-465809

RESUMO

OBJETIVOS: Avaliar a função e a qualidade de vida dos pacientes pós-artroplastia de Girdlestone e comparar os resultados entre os grupos Girdlestone unilateral e o grupo com prótese total de quadril contralateral. MÉTODOS: estudo transversal no qual foram avaliados 9 pacientes com Girdlestone unilateral e 3 com Girdlestone em um quadril e prótese total no quadril contralateral. A avaliação constitui-se em aplicar o questionário genérico de qualidade de vida SF-36 e um questionário funcional específico para o quadril, Harris Hip Score (HHS). A comparação dos grupos foi realizada usando-se o teste t- Student e o teste de Fisher. RESULTADOS: Os pacientes do grupo Girdlestone unilateral apresentaram maior quantidade de domínios do SF-36 classificados como elevados, embora 77,8 por cento destes tenham obtido resultados ruins no HHS. Todos os pacientes apresentaram o teste de Trendelenburg positivo e discrepância de membros, o que levou à marcha claudicante em 11 dos 12 pacientes avaliados. Destes, apenas 6 submeteram-se a fisioterapia pós-operatória. CONCLUSÃO: A qualidade de vida e a função pós-operatória de Girdlestone, na população brasileira, ainda necessita ser mais pesquisada, pois estes resultados são indicações do comportamento das variáveis de estudo e não podem ser consideradas encerradas.


OBJECTIVES: To evaluate function and quality of life of patients submitted to Girdlestone's arthroplasty, and to compare outcomes between unilateral Girdlestone's group with the group with contralateral total hip prosthesis. METHODS: Cross-sectional study where 9 patients were evaluated with unilateral Girdlestone's and 3 with Girdlestone's in one hip and contralateral total hip prosthesis. The evaluation consisted in filling in a generic questionnaire on quality of life "SF-36" and a specific questionnaire for hip function "Harris Hip Score" (HHS). The comparison between groups was made by using the Student's t-test and the Fisher's test. RESULTS: The patients of the unilateral Girdlestone's group presented a higher number of SF-36 domains classified as high, although 77.8 percent of these showed poor results on the HHS. All patients had a leg-length discrepancy and positive Trendelenburg's test, which led to limping gait in 11 of 12 patients evaluated. Of these, only 6 underwent physiotherapy after surgery. CONCLUSION: Girdlestone's postoperative quality of life and function in a Brazilian population still requires further studies, because these outcomes are indicative of study variables' behavior and cannot be regarded as definite.


Assuntos
Humanos , Masculino , Feminino , Adulto , Pessoa de Meia-Idade , Idoso de 80 Anos ou mais , Artroplastia , Artroplastia de Quadril , Estudos Transversais , Qualidade de Vida , Inquéritos e Questionários
12.
Acta ortop. bras ; 13(2)2005. ilus
Artigo em Português, Inglês | LILACS | ID: lil-404150

RESUMO

Apresentamos o relato de um caso de osteocondroma solitário acetabular, tumor ósseo comum em diversas regiões do esqueleto porém raro nesta localização. O tratamento realizado foi a ressecção do tumor por uma via posterior ao quadril e o paciente evoluiu com resolução do quadro clínico.


Assuntos
Humanos , Feminino , Adolescente , Acetábulo/fisiopatologia , Neoplasias Ósseas , Osteocondroma , Osteocondroma , Acetábulo/cirurgia , Cabeça do Fêmur , Quadril , Osteotomia/reabilitação
13.
Rev. bras. reumatol ; 44(3): 185-191, maio-jun. 2004. ilus, tab
Artigo em Português | LILACS | ID: lil-392024

RESUMO

Objetivo: avaliar a capacidade funcional da articulação do quadril e a qualidade de vida de pacientes portadores de osteoartrose submetidos à artroplastia total. Métodos: foram avaliados 12 pacientes através do questionário Harris Hip Score (modificado/escala de 9-95 pontos) e goniometria no pré e pós-operatório de artroplastia total de quadril. Resultados: a amplitude de movimento mostrou-se incompleta em todos os arcos de movimento no pré-operatório. No pós-operatório os valores aumentaram significativamente na flexão (p=0,007), extensão (p=0,02), abdução (p=0,001), rotação medial (p=0,0002) e lateral (p=0,01), apenas o aumento do movimento de adução de quadril não foi significativo (p=0,08). A medida do escore obtido através do questionário no pré-operatório foi 39,7 pontos e no pós-operatório 74,1 pontos (P<0,000001). Conclusões: a artroplastia total do quadril proporciona ao paciente alívio da dor, melhora na realização das atividades da vida diária e aumento da amplitude do movimento articular do quadril. Os pacientes, porém, relatam medo e insegurança na realização dessas atividades.


Assuntos
Humanos , Masculino , Feminino , Adulto , Pessoa de Meia-Idade , Artroplastia , Quadril , Osteoartrite , Qualidade de Vida , Inquéritos e Questionários
14.
Rev. bras. ortop ; 36(5): 149-154, maio 2001. ilus, tab
Artigo em Português | LILACS | ID: lil-334916

RESUMO

Este trabalho apresenta os resultados obtidos na simulação computadorizada da artroplastia total do quadril. Estuda o desgaste do polietileno do componente acetabular, utilizando o método de elementos finitos, com o objetivo de verificar a validade deste método de simulação. Foi simulado o desgaste do polietileno produzido pelo contato da cabeça femoral de cromo-cobalto, tendo o componente acetabular orientação espacial de 45° de inclinação lateral e anteversão neutra. Os resultados encontrados foram semelhantes àqueles observados em estudos clínicos.


Assuntos
Análise de Elementos Finitos , Artroplastia de Quadril , Polietileno
15.
Cir. vasc. angiol ; 16(3): 108-113, jun. 2000. tab
Artigo em Português | LILACS | ID: lil-301469

RESUMO

O tromboembolismo venoso (TEV, trombose venosa profunda e embolia pulmonar) é uma séria complicaçäo dos procedimentos cirúrgicos de grande porte, especialmente nas cirurgias de quadril e joelho.Com o objetivo de avaliar o uso da profilaxia medicamentosa para TVP, bem como o índice de suspeita e o manuseio desta complicaçäo nos pacientes submetidos à prótese total de quadril, realizou-se estudo retrospectivo baseado nos prontuáarios de pacientes de 16 hospitais brasileiros, 7 públicos e 9 privados, durante os anos de 1997 e 1998.Os centros de ortopedia convidados a participar preencheram um formulário especificando sexo, idade, data da cirurgia, principal indicaçäo para prótese de quadril, doenças concomitantes e o uso de prodilaxia medicamentosa, incluindo o tipo de droga e a posologia.Outros dados coletados foram volume de transfusäo sanguínea, suspeita de TVP até 15 dias após a alta e o manuseio da TVP após a suspeita.Resultados:dados de 386 pacientes de 16 hospitais foram coletados; 310 (80 por cento) receberam algum tipo de profilaxia medicamentosa, 20 por cento nenhuma.Dos que receberam profilaxia, heparina näo fracionada (HNF) foi a droga mais comumente usada, em 55 (14 por cento) dos pacientes.Seu uso combinado ou associado à aspirina foi encontrado em 7 (2 por cento) dos pacientes.Nos 248 pacientes em que HFN foi prescrita, 17 (7por cento) receberam 5000 UI 3x/dia, e 93 por cento receberam 2x/dia...


Assuntos
Humanos , Feminino , Adulto , Embolia Pulmonar , Tromboembolia , Prótese de Quadril , Incidência , Estudos Prospectivos , Resultado do Tratamento
16.
Rev. bras. ortop ; 34(2): 145-154, fev. 1999. ilus, tab
Artigo em Português | LILACS | ID: lil-360900

RESUMO

Os autores avaliam, em estudo prospectivo cl¡nico e radiográfico, a haste de Spotorno em 47 artroplastias totais do quadril não-cimentadas, realizadas em 40 pacientes. Todos os procedimentos foram realizados por via lateral e mesma técnica cir£rgica. Ap¢s seguimento médio de seis anos e um mês, m¡nimo de cinco anos e três meses e máximo de sete anos, apresentam-se os resultados, em que a dor não foi relatada como importante por nenhum paciente; em 43 (91,5 por cento) artroplastias, os pacientes estavam satisfeitos com o resultado da cirurgia; não se observou qualquer sinal de osteopenia em 11 (23,4 por cento) exames radiográficos e não houve formação de pedestal em 36 (76,6 por cento) fêmures. De acordo com a metodologia empregada, 8 (17,0 por cento) componentes femorais apresentavam sinais sugestivos de soltura e 39 (83 por cento) das 47 artroplastias foram consideradas satisfat¢rias.


Assuntos
Humanos , Masculino , Feminino , Adulto , Pessoa de Meia-Idade , Artroplastia de Quadril , Fêmur
17.
Rev. bras. ortop ; 34(1): 55-58, jan. 1999. ilus, graf
Artigo em Português | LILACS | ID: lil-360884

RESUMO

Os autores apresentam os resultados obtidos nas artroplastias totais do quadril em pacientes portadores de protusão acetabular, utilizando a cabeça femoral como enxerto aut¢geno, não delaminado, e, em forma semi-esférica, com o objetivo de reforçar a parede acetabular medial. Foram operados 16 pacientes (19 quadris) com diagn¢stico de osteoartrose (sete), artrite reumat¢ide (seis), l£pus eritematoso sistêmico (um), artrite psoriática (um) e seqüela de pioartrite (um); oito (42,10 por cento) quadris apresentavam protusães de grau III, nove (47,37 por cento) de grau II e dois (10,53 por cento) de grau I. Os resultados foram considerados satisfat¢rios em todos os pacientes, de acordo com o método de avaliação de Merle dÆAubigné (1970).


Assuntos
Humanos , Masculino , Feminino , Adulto , Pessoa de Meia-Idade , Artroplastia de Quadril
18.
Rev. bras. ortop ; 30(7): 443-51, jul. 1995. ilus, tab
Artigo em Português | LILACS | ID: lil-161113

RESUMO

Os autores apresentam uma via de acesso lateral do quadril com osteotomia parcial do trocanter maior, no intuido de preservar a integridade da musculatura abdutora do quadril, evitando as maceraçöes e as desnervaçöes. Foram incluídos neste estudo 31 pacientes, 31 quadris, submetidos à artroplastia total do quadril superior pela abordagem acima proposta e avaliadas as complicaçöes intra e pós-operatórias imediatas, assim como o estudo da força abdutora do quadril e a consolidaçäo da osteotomia. Os pacientes näo apresentaram complicaçöes na consolidaçäo do fragmento osteotomizado; as fraturas do trocanter maior, ocorridas em dois (6,5 porcento) pacientes, näo foram atribuídas à via de acesso. Com relaçäo ao resultado funcional, apenas um (3,2 porcento) paciente apresentou diminuiçäo da força de abduçäo do quadril. Os autores recomendam a utilizaçäo desta via de acesso especialmente nas artropatias primárias


Assuntos
Humanos , Masculino , Feminino , Pessoa de Meia-Idade , Adulto , Fêmur/cirurgia , Osteotomia , Prótese de Quadril/métodos , Articulação do Quadril , Complicações Intraoperatórias , Complicações Pós-Operatórias , Resultado do Tratamento
19.
Rev. bras. ortop ; 30(7): 473-80, jul. 1995. tab
Artigo em Português | LILACS | ID: lil-161119

RESUMO

Os autores estudaram o sintoma de dor na coxa, pós-artroplastia total do quadril, com prótese näo cimentada tipo PCA, utilizada em 79 pacientes operados no período de agosto de 1987 a setembro de 1989. Correlacionaram-na ao surgimento deste sintoma com possíveis instabilidades do implante femoral utilizado


Assuntos
Humanos , Masculino , Feminino , Adulto , Pessoa de Meia-Idade , Medição da Dor , Falha de Prótese , Prótese de Quadril/efeitos adversos , Seguimentos , Coxa da Perna , Resultado do Tratamento
20.
Folha méd ; 109(2): 69-71, ago. 1994.
Artigo em Português | LILACS | ID: lil-159174

RESUMO

Os autores revisaram e compilaram os trabalhos referentes a artroplastias do quadril desde os mais antigos autores e os mais recentes, com vistas a tratados atuais e as referências históricas do assunto. Incluem a experiência pessoal em relaçåo à utilizaçåo das mais modernas condutas e as mais recentes contribuiçÆes à artroplastia do quadril. Enfim, é um relato histórico e atual que elucida aspectos da evoluçåo do assunto


Assuntos
Humanos , Prótese de Quadril/história , Prótese de Quadril/métodos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA